2. DataMatrix Code Specifications for Medical Devices

Shenzhen, China
Published: Jul 3, 2026
2. DataMatrix Code Specifications for Medical Devices

The 2D DataMatrix (ECC 200) is the standard AIDC format for medical device UDI marking.

Code Size and Cell Specifications

Parameter Standard Application Small Instruments
Code size 2.5–4.5 mm² As small as 0.4 mm²
Cell size (x-dimension) ~0.300 mm ~0.200 mm
Quiet zone Per ISO/IEC 16022 Per ISO/IEC 16022
Error correction ECC 200 (Reed-Solomon) ECC 200 (Reed-Solomon)
Quality grade Per ISO/IEC 15415 Per ISO/IEC 15415

Practical Sizing Examples

Device Mark Size Mark Content Marking Time
Surgical scissors (SS316) 2.71×2.71mm Logo + 2D DataMatrix ~5 seconds
Bone screw (Ti, 1–2mm dia.) ~0.4mm Micro DataMatrix
Implant tray 4.5×4.5mm Full UDI + text ~8 seconds
HDPE bottle 2.71mm² Logo + 2D DataMatrix ~2 seconds